Feature
Conductivity and TDS measurement. Autoranging & temperature compensation for higher accuracy.
Overview
Determine water conductivity with as little as 0.12 mL of sample.
The conductivity of rain water is a trusted guide to determining atmospheric purity. In agriculture, measuring the conductivity of soil allows farmers and agronomists to determine optimum fertiliser usage and check the ‘health’ of soil after salt water damage. The LAQUAtwin meter makes conductivity testing simple, anywhere.
Applications include:
Fresh water testing (rain, rivers, lakes, hot springs); aquaria; soil testing for improved agriculture and analysis of salt water damage; analysis of surface cleanliness prior to coating; and improved paint adhesion.
Functions
- Salt/TDS Measurement
- Auto range change
- Temperature conversion(2%/℃ fixed)
- IP67 waterproof
- Auto hold
- Automatic power off(15 minutes)
| Model |
EC 11 |
EC 22 |
EC 33 |
| Measurement principle |
2 Electrode bipolar AC Titanium coated with Platinum black |
||
| Minimum sample volume |
0.12 mL |
||
| Measurement range |
Conductivity 0 to 199 uS/cm 200 to 1999 uS/cm 2.00 to 19.99 mS/cm |
Conductivity 0 to 199 uS/cm 200 to 1999 uS/cm 2.00 to 19.99 mS/cm 20.0 to 199.9 mS/cm |
|
|
— |
TDS 0.0 to 99.9 ppm 100 to 999 ppm 1000 to 9990 ppm |
||
| Resolution |
Conductivity 1 uS/cm 1 uS/cm 0.01 mS/cm |
Conductivity 1 uS/cm 1 us/cm 0.01 mS/cm 0.1 mS/cm |
|
|
— |
TDS 0.1 ppm; 1 ppm; 10 ppm |
||
| Calibration |
Two – point |
Three – point |
|
| Accuracy |
2 % F.S. 1 digit (for each range) |
||
| Calibration curves |
TDS Factor (0.4 to 1.0) / EN 27888 / 442 / NaCI |
||
| Functions |
Temperature compensation , IP67 Water Proof , Auto Hold , Auto Stable , Automatic power off (30 minutes) |
||
| Display |
Custom digital LCD |
||
| Operating temperature / Humidity |
5 to 40 C, 85% or less in relative humidity (No condensation) |
||
| Battery life |
Approx. 400 hours in continuous use with x2 CR 2032 batteries |
||
| Main Material |
ABS epoxy |
||
| Dimensions / Mass |
164 mm x 29 mm x 20 mm (excluding projections) / Approx. 50 g (meter only, without batteries, approx. 45g) |
||
| Accessories included |
14 mL Standard solutions (1413 uS/cm & 12.88 mS/cm) and 4 mL Conditioning solution |
||